HOMECHOICE
Report
- Report Number
- 1416980-2012-06745
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- September 29, 2012
- Report Date
- November 21, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Removal / Correction Number
- 1423500-01/08/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012 22:19:06. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1519ML, INDICATING THE HOME PATIENT (HP) DRAINED 1519ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2400ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |