FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 2862476 · Received December 10, 2012

Report

Report Number
3008382007-2012-07500
Event Type
Malfunction
Date Received
December 10, 2012
Report Date
November 29, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. THE OTHER PRODUCTS HAVE ALSO BEEN RETURNED BUT NOT YET EVALUATED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4): THE TEST STRIP HAS PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH PING METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON THE MORNING OF (B)(6) 2012. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "230 MG/DL" WITH THE SUBJECT METER AND "134 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF (B)(4). THE PATIENT MANAGES HIS DIABETES WITH INSULIN THROUGH PUMP THERAPY. BASED ON THE ALLEGED RESULT, THE PATIENT REPORTEDLY ADMINISTERED SELF NOVOLOG INSULIN (3.8 UNITS). THE PATIENT DENIED DEVELOPING SYMPTOMS DUE TO THE REPORTED ISSUE. NO ADDITIONAL TREATMENT WAS SPECIFIED. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THERE IS NO INDICATION THAT THE REPORTED ISSUE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY SINCE THE PATIENT DENIED DEVELOPING SYMPTOMS. THERE IS NO EVIDENCE THE PATIENT ADMINISTERED INAPPROPRIATE SELF TREATMENT DUE TO THE ALLEGED ISSUE. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE SUBJECT METER DID NOT MEET LFS'S ACCURACY CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3298244

Patients

Seq Age Sex Outcome Treatment
1 55 YR