ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-14319
- Event Type
- Injury
- Date Received
- December 10, 2012
- Date of Event
- November 8, 2012
- Report Date
- November 8, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A RESERVED SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED AND NO DAMAGE OR DEFECTS WERE NOTED. A FILL, FORCE AND LEAK TEST WAS PERFORMED WITH NO FAILURES. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS ALLEGING THE FOLLOWING: THE CARTRIDGES (LOT # B201865 EXP DATE 05/2014 ) ARE DAMAGED PRIOR TO USE AND THE PATIENT IS UNABLE TO SECURE THE TUBING TO THE CARTRIDGE PROPERLY, CAUSING THE TUBING TO CONNECT ON CROOKEDLY. ALLEGEDLY DUE TO THE ISSUE THE PATIENT HAS HAD BLOOD GLUCOSE (BG)S AS HIGH AS THE 400'S MG/DL WITH SYMPTOMS OF KETONES IN URINE, AND FATIGUE, AND AS LOW AS 61 MG/DL (NO SYMPTOMS). THE PATIENT HAS CHANGED HIS SITE MULTIPLE TIMES AND NOTICED NO CHANGE IN THE BG. PARENT CHANGED THE CARTRIDGE 3-4 TIMES (PARENT IS A POOR HISTORIAN) AND WITH THE LAST CARTRIDGE CHANGE SHE HAS NOTICED THAT THE BG'S ARE IN NORMAL RANGE. THE REPORTER STATES THE CARTRIDGES ARE USED PER INSTRUCTIONS, AND THE PATIENT IS CONTINUING TO SES AIR BUBBLES 2-3 INCHES IN THE TUBING THE PUMP NOT AVAILABLE FOR REVIEW TO CONFIRM PATIENT DISCONNECTION. THE MOST RECENT BG WAS 114MG/DL. PATIENT HAS RESUMED USE OF THE PUMP AT THIS TIME USING A CARTRIDGE FROM A DIFFERENT BOX. CUSTOMER SUPPORT ADVISED THE PARENT NOT TO USED DEFECTIVE CARTRIDGES, TO MONITOR THE BG, AND CALL BACK WITH ANY OTHER ISSUES. THIS COMPLAINT IS BEING REPORTED BASED ON THE ALLEGATION THAT DEFECTIVE CARTRIDGES MAY HAVE LEAD TO DISRUPTION IN INSULIN DELIVERY AND HYPERGLYCEMIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS IR 1000 CARTRIDGE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | Life Threatening |