FDA Adverse Event Malfunction Summary report: N

PROXIMATE PPH PROCEDURE SET

MDR report key: 2862388 · Received December 10, 2012

Report

Report Number
3005075853-2012-05514
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 5, 2012
Report Date
November 15, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051301
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE CONTACT. TWO DEVICES WERE RETURNED FOR ANALYSIS: DEVICE (A) ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. DEVICE (B) ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN INTENDED PROLAPSECTOMY ACCORDING TO THE LONG PROCEDURE, THE FIRST DEVICE USED CREATED ONLY A POSTERIOR HALF STAPLE LINE. THEREFORE, A SECOND DEVICE WAS USED TO RESECT THE MISSING ANTERIOR PART. AFTER EXTRACTING THE SECOND DEVICE, THE SURGEON NOTICED AN ALMOST COMPLETE DIASTASES OF THE ANASTOMOSIS. THE SURGEON HAD TO APPLY INTERRUPTED SUTURES ON THE ANASTOMOTIC STAPLE LINE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE PPH PROCEDURE SET STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4CG9J

Patients

Seq Age Sex Outcome Treatment
1 45 YR