FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2862097 · Received November 29, 2012

Report

Report Number
1720753-2012-09829
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
November 18, 2012
Report Date
November 29, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS RESEATED AND THE CABLES WERE RECONNECTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT PRODUCE X-RAY AFTER THEY MOVED IT TO ANOTHER PLACE. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1