VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2012-00897
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Date of Event
- November 7, 2012
- Report Date
- November 9, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INSPECTION AND CONNECTED TO A WATER BAG TO CHECK FOR LEAKS. RESULTS: UPON CONNECTING TO THE WATERBAG, A DROP OF WATER BEGAN TO BUILD AT THE CONNECTION BETWEEN THE FEEDSET TUBE AND THE WATERBAG SPIKE. A SUFFICIENT AMOUNT OF GLUE WAS FOUND AT THE SPIKE AND FEEDSET TUBE CONNECTION. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 120501. CONCLUSION: A SUFFICIENT AMOUNT OF GLUE WAS APPLIED TO THE SPIKE TUBING CONNECTIONS, BUT THE GLUE WAS FOUND TO BE PARTLY BONDING. WE WERE UNABLE TO DETERMINE THE CAUSE OF THE FAILURE. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS PRODUCT WOULD HAVE FAILED THE FINAL PRESSURE TEST IF IT WAS IN A BROKEN STATE DURING TESTING. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE FEEDSET LINE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS DAMAGED DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 120501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |