SPECTRUM INFUSION PUMP
MDR report key: 2861860
·
Received November 8, 2012
Report
- Report Number
- 1314492-2012-00402
- Event Type
- Malfunction
- Date Received
- November 8, 2012
- Report Date
- October 10, 2012
- Manufacturer
- BAXTER HEALTHCARE CORP.
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE SIGMA DEVICE EVALUATION FOUND THE #5, #6, #7, #8, AND #9 KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT ON THE #6 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED, 16 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AP WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE #4, #5 AND #6 KEYS ON A PUMP KEYPAD ARE INOPERABLE.