FDA Adverse Event Summary report: N

BRILLIANCE 6 AIR

MDR report key: 2861692 · Received November 28, 2012

Report

Report Number
1525965-2012-00109
Date Received
November 28, 2012
Report Date
October 31, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K012009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED AFTER A PATIENT SCAN WAS COMPLETED AND DURING VERTICAL DOWN MOVEMENT OF THE PATIENT TABLE, THE PATIENT TABLE DROPPED APPROXIMATELY 2 INCHES AND THEN STOPPED. THE PATIENT WAS JARRED AND STARTLED, BUT WAS NOT HARMED. THERE IS POTENTIAL FOR SERIOUS INJURY TO A PATIENT OR OPERATOR IF THIS WERE TO RECUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BRILLIANCE 6 AIR JAK PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. 728256

Patients

Seq Age Sex Outcome Treatment
1