FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNEL

MDR report key: 2861581 · Received November 28, 2012

Report

Report Number
9615050-2012-01422
Event Type
Malfunction
Date Received
November 28, 2012
Date of Event
October 22, 2012
Report Date
October 30, 2012
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND AT POWER UP, THE DEVICE DISPLAYED A WHITESCREEN ERROR AND THE DEVICE SOUNDED AN AUDIBLE ALARM FROM THE SECONDARY BEEPER. THIS WAS DUE TO THE HARDWARE MODULE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED A WHITESCREEN ERROR. THE PUMP WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT STATED, "PUMP NOT WORKING." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY AFTER THE PUMP WAS POWERED ON, THE PUMP ALARMED WITH A WHITESCREEN ERROR. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYMBIQ SINGLE CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK