FLOGARD
Report
- Report Number
- 1416980-2012-06644
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Date of Event
- November 9, 2012
- Report Date
- November 15, 2012
- Manufacturer
- BAXTER HEALTHCARE - BUFFALO GROVE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A 38 FAILURE CODE WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS IDENTIFIED AS AN UNPLUGGED FORCE SENSING RESISTORS (FSRS) - RIBBON CABLE. THE FSRS AND RIBBON CABLE WERE REPLACED TO CORRECT THE REPORTED CONDITION. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A 38 FAILURE CODE. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - BUFFALO GROVE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |