FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2861578 · Received December 7, 2012

Report

Report Number
1416980-2012-06644
Event Type
Malfunction
Date Received
December 7, 2012
Date of Event
November 9, 2012
Report Date
November 15, 2012
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A 38 FAILURE CODE WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS IDENTIFIED AS AN UNPLUGGED FORCE SENSING RESISTORS (FSRS) - RIBBON CABLE. THE FSRS AND RIBBON CABLE WERE REPLACED TO CORRECT THE REPORTED CONDITION. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A 38 FAILURE CODE. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - BUFFALO GROVE

Patients

Seq Age Sex Outcome Treatment
1