PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-07710
- Event Type
- Injury
- Date Received
- December 7, 2012
- Date of Event
- November 19, 2012
- Report Date
- November 19, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE ANALYSIS OF THE RETURNED DEVICE REVEALED THE POSTERIOR CUFF WAS MISSED AND THIS WOULD RESULT IN A FAILURE TO RETRIEVE THE SUTURE DURING THE NEEDLE PLUNGER RETRACTION AND MAY APPEAR SIMILAR TO A SUTURE BREAK. THEREFORE, THE REPORTED SUTURE BREAK WAS NOT CONFIRMED. BASED ON THE REPORTED INFORMATION, MANUFACTURING INSPECTION CRITERIA AND ANALYSIS OF THE RETURNED DEVICE, PROBABLE CAUSE FOR THE REPORTED EVENT WAS DETERMINED TO BE RELATED TO THE OPERATIONAL CONTEXT DURING USE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE WITH A 6FR SHEATH, AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A SUTURE BREAK OCCURRED. A NON-ABBOTT DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 21003J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention | SHEATH: 6FRHEPARIN |