FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2861195 · Received November 2, 2012

Report

Report Number
1314492-2012-00389
Event Type
Malfunction
Date Received
November 2, 2012
Report Date
October 3, 2012
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIGMA EVALUATED THE DEVICE IN QUESTION. IT WAS CONFIRMED THAT THE AUDIO FUNCTION WAS NOT PRESENT. FURTHER EVALUATION IDENTIFIED THAT THE ABSENCE OF AUDIO WAS DUE TO FAILED 1 WATT SPEAKER. ALL DETECTION CAPABILITIES AND VISUAL FUNCTIONS PERFORMED AS EXPECTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP DID NOT HAVE AUDIO FUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1