OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2012-00550
- Event Type
- Malfunction
- Date Received
- November 2, 2012
- Date of Event
- October 4, 2012
- Report Date
- October 4, 2012
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR ASSESS WHETHER IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER'S GUIDE WARNS "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER."
THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE WAS HIGH, MEASURING 269, 256, AND 211 MG/DL (NO TIMES OR TREATMENT REPORTED) AND WHEN THE DEVICE WAS REMOVED THE CANNULA WAS BENT AT A 90 DEGREE ANGLE.