FDA Adverse Event Malfunction Summary report: N

HX2 TEMPERATURE MANAGEMENT SYSTEM

MDR report key: 2860771 · Received December 5, 2012

Report

Report Number
1828100-2012-01566
Event Type
Malfunction
Date Received
December 5, 2012
Date of Event
November 13, 2012
Report Date
November 13, 2012
Manufacturer
TERUMO CARDIOVASCULAR SYSTEM CORP
Product Code
DWC
PMA / PMN Number
K071521
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REPRESENTATIVE (FSR). THE FSR REPLACED THE RESISTANCE TEMPERATURE DETECTOR (RTD) CABLE ASSEMBLY, THE DUAL THERMISTOR, THE D/C BOARD ON CHANNEL A, ALONG WITH A LONGER CIRCULATING LOOP. WITH THE PARTS REPLACED, THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THAT AN "EO4" ERROR MESSAGE OCCURRED ON CHANNEL A. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HX2 TEMPERATURE MANAGEMENT SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEM CORP 809810

Patients

Seq Age Sex Outcome Treatment
1