FDA Adverse Event Malfunction Summary report: N

EXCELCARE BARIATRIC BED FRAME

MDR report key: 2860746 · Received October 30, 2012

Report

Report Number
1045510-2012-00032
Event Type
Malfunction
Date Received
October 30, 2012
Date of Event
October 23, 2012
Report Date
October 23, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

HILL-ROM RECEIVED A COMPLAINT ON ONE OF ITS EXCELCARE BARIATRIC BED FRAMES ALLEGING THE CPR ASSEMBLY WAS WORKING INTERMITTENTLY. A HILL-ROM SERVICE TECHNICIAN PERFORMED AN INVESTIGATION AND FOUND THE CPR HANDLE BROKEN DURING HIS INSPECTION. HE REPLACED THE HANDLE, TIGHTENED THE CABLE AND ADJUSTED IT TO RETURN FUNCTION TO THE CPR ASSEMBLY. HILL-ROM IS REPORTING THIS MALFUNCTION IN COMPLIANCE WITH 21CFR803.3 WHERE THIS FAILURE MODE WAS INVOLVED IN AN ADVERSE EVENT ON (B)(4) 2009 INDICATED IN MDR 1045510-2009-0021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXCELCARE BARIATRIC BED FRAME AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 90812

Patients

Seq Age Sex Outcome Treatment
1