HOMECHOICE PRO
Report
- Report Number
- 1416980-2012-06624
- Event Type
- Malfunction
- Date Received
- December 6, 2012
- Date of Event
- October 14, 2011
- Report Date
- November 12, 2012
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Removal / Correction Number
- 1423500-01/08/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
A HOMECHOICE UNIT (SERIAL NUMBER: (B)(4)) WAS RETURNED TO BAXTER FOR A SOFTWARE UPGRADE, DUE TO A FIELD CORRECTIVE ACTION (2009FCA07) INITIATED (B)(6) 2011 BY BAXTER (B)(4). IN ADDITION TO THE SOFTWARE UPGRADE, AN EVALUATION OF THE DEVICE EVENT HISTORY LOG WAS CONDUCTED ON (B)(6) 2012. AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED, WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2011 AT 22:49:41H. AT THIS TIME BAXTER WAS NOT CONTACTED BY THE PATIENT FOR ASSISTANCE. THE DETAILS OF THE IIPV EVENT WERE AS FOLLOWS; DURING NIGHT DRAIN CYCLE SEVEN, THE PATIENT'S ULTRAFILTRATION READING WAS 1604ML, INDICATING THE HOME PATIENT (HP) DRAINED 1604ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA, AND IS A REPORTABLE MALFUNCTION. IT IS UNKNOWN HOW LONG THE DEVICE WAS IN USE WHEN THE EVENT OCCURRED. FURTHERMORE, AS THIS EVENT WAS FOUND DURING SERVICE, ANY PATIENT INJURY OR MEDICAL INTERVENTION IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |