FDA Adverse Event
Malfunction
Summary report: N
ENTERALITE INFINITY ENTERAL FEEDING PUMP
MDR report key: 2860446
·
Received November 5, 2012
Report
- Report Number
- 1722139-2012-01124
- Event Type
- Malfunction
- Date Received
- November 5, 2012
- Date of Event
- October 1, 2012
- Report Date
- October 31, 2012
- Manufacturer
- MOOG MEDICAL DEVICES GROUP
- Product Code
- LZH
- PMA / PMN Number
- K031199
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB WAS REPLACED TO SOLVE THE ISSUE.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THAT PUMP WAS PART OF IDENTIFIED POPULATION AT RISK AND FAILED TH DIODE TEST DURING TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENTERALITE INFINITY ENTERAL FEEDING PUMP | LZH | MOOG MEDICAL DEVICES GROUP | INFINITY PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |