FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2860352
·
Received November 5, 2012
Report
- Report Number
- 1824206-2012-07112
- Event Type
- Malfunction
- Date Received
- November 5, 2012
- Date of Event
- October 16, 2012
- Report Date
- October 16, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECH EXAMINED THE POWER CORD AND FOUND A LOOSE GROUND PIN. HE REPLACED THE POWER CORD AND THE BED FUNCTIONED TO SPECIFICATIONS.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THE BED'S GROUND RESISTANCE WAS OUT OF SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |