FDA Adverse Event Other Summary report: N

ENDOTOOL

MDR report key: 2860283 · Received November 26, 2012

Report

Report Number
3009864844-2012-00002
Event Type
Other
Date Received
November 26, 2012
Date of Event
October 31, 2012
Report Date
November 26, 2012
Manufacturer
ENDOTOOL LLC
Product Code
NDC
PMA / PMN Number
K053137
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

BOTH PTS WERE JUST STARTED ON ENDOTOOL USING THE "START FROM IV INSULIN" RADIO BUTTON. THE RN CALLED ENDOTOOL WHEN HE/SHE RECEIVED NO DOSE. I INSTRUCTED THEM TO DELETE THE READING IN ENDOTOOL, REPEAT THE GLUCOSE AND THEN SELECT "ENDOTOOL TO CALCULATE¿" RADIO BUTTON. THE NURSE DID SO AND PUT IN A BG AND RECEIVED A DOSE OF INSULIN FOR THE PT.

Additional Manufacturer Narrative · 1

BOTH PTS WERE JUST STARTED ON ENDOTOOL USING THE "START FROM IV INSULIN" RADIO BUTTON. THE RN CALLED ENDOTOOL WHEN HE/SHE RECEIVED NO DOSE. I INSTRUCTED THEM TO DELETE THE READING IN ENDOTOOL, REPEAT THE GLUCOSE AND THEN SELECT "ENDOTOOL TO CALCULATE¿" RADIO BUTTON. THE NURSE DID SO AND PUT IN A BG AND RECEIVED A DOSE OF INSULIN FOR THE PT.

Description of Event or Problem · 1

TWO EVENTS OF POSSIBLE MALFUNCTION OF THE ENDOTOOL DRUG DOSE CALCULATOR OCCURRED ON THE DATES STATED ABOVE AT THE SAME HOSP. THE FIRST INCIDENT THE ENDOTOOL SYS DID NOT CALCULATE A DOSE FOR THE PT FOR ALMOST TWO DAYS WITH A BLOOD GLUCOSE LEVEL OF 170. IN THE SECOND INCIDENT A PT DID NOT GET A DOSE CALCULATED OR RECEIVE A DOSE OF INSULIN WITH A BLOOD GLUCOSE OF 160. THE NURSE WAS CALLED AND ENDOTOOL WAS CONTACTED. THE CLINICAL COORDINATE INSTRUCTED THE NURSE TO DELETE THE READING IN ENDOTOOL, REPEAT THE GLUCOSE AND TO PRESS THE CALCULATE RATION BUTTON. THE PT WAS THEN DOSED. A 45 MINUTE DELAY IN DOSING OCCURRED. NO HYPER OR HYPO-GLYCEMIA OCCURRED IN EITHER PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTOOL BLOOD GLUCOSE DOSE CALCULATOR NDC ENDOTOOL LLC VERSION 2.27

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention