FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNEL

MDR report key: 2860216 · Received November 21, 2012

Report

Report Number
9615050-2012-01401
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 16, 2012
Report Date
October 17, 2012
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT PRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING VERIFICATION TESTING AT THE SERVICE CENTER, WHEN PLUGGED INTO AC POWER, SPARKS AND CHARRING WERE NOTED AT THE AC RECEPTACLE AND THE POWER CORD OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYMBIQ SINGLE CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA