FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2860064 · Received November 21, 2012

Report

Report Number
8020893-2012-01264
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 4, 2012
Report Date
November 6, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI BACKLIGHT INVERTER PCB. THE UNIT PASSED EXTENDED SELF-TESTING.