ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT
Report
- Report Number
- 1419322-2012-00040
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Report Date
- October 10, 2012
- Manufacturer
- DENTSPLY PROFESSIONAL
- Product Code
- EKX
- PMA / PMN Number
- K031145
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- DENTIST
Narratives
THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE BUR END BEARING FAILURE, FREE ROAMING BALL BEARINGS AND EXCESSIVE DEBRIS MOST LIKELY CREATED SUBSTANTIAL FRICTION WITHIN THE HEAD WHICH RESULTED IN TEMPERATURE INCREASE. IT IS REASONABLE TO SUSPECT GEAR FUNCTION WAS COMPROMISED BY THE ADDED DEBRIS INSIDE THE HEAD WHICH IS EVIDENT FROM THE ABRASIVE WEAR ON THE TEETH AND OUTER SURFACE OF THE GEARS. A LACK OF LUBRICATION WAS ALSO NOTICED AS RECEIVED.
IN THIS EVENT, A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT | EKX | DENTSPLY PROFESSIONAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |