FDA Adverse Event Malfunction Summary report: N

ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT

MDR report key: 2860042 · Received October 31, 2012

Report

Report Number
1419322-2012-00040
Event Type
Malfunction
Date Received
October 31, 2012
Report Date
October 10, 2012
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
EKX
PMA / PMN Number
K031145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE BUR END BEARING FAILURE, FREE ROAMING BALL BEARINGS AND EXCESSIVE DEBRIS MOST LIKELY CREATED SUBSTANTIAL FRICTION WITHIN THE HEAD WHICH RESULTED IN TEMPERATURE INCREASE. IT IS REASONABLE TO SUSPECT GEAR FUNCTION WAS COMPROMISED BY THE ADDED DEBRIS INSIDE THE HEAD WHICH IS EVIDENT FROM THE ABRASIVE WEAR ON THE TEETH AND OUTER SURFACE OF THE GEARS. A LACK OF LUBRICATION WAS ALSO NOTICED AS RECEIVED.

Description of Event or Problem · 1

IN THIS EVENT, A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT EKX DENTSPLY PROFESSIONAL

Patients

Seq Age Sex Outcome Treatment
1