FDA Adverse Event Injury Summary report: N

PINNACLE 100 ACET CUP 54MM

MDR report key: 2859700 · Received December 6, 2012

Report

Report Number
1818910-2012-83196
Event Type
Injury
Date Received
December 6, 2012
Date of Event
October 16, 2012
Report Date
May 31, 2013
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LPH
PMA / PMN Number
PK000306
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL. THERE IS NO NEW INFORMATION THAT WOULD CHANGE THE EXISTING MDR DECISION. RECORDS ARE AVAILABLE FOR FURTHER REVIEW.DEPUY CONSIDERS THIS COMPLAINT CLOSED AT THIS TIME.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORD FOR PRODUCT 121887354 LOT 2210707 REVEALED NO RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. D. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED

Description of Event or Problem · 1

CLINICAL REPORT STATES THE PATIENT WAS REVISED DUE TO METALLOSIS AND INCREASING HIP PAIN. MEDICAL RECORDS INDICATE THE PATIENT HAD MODERATELY ELEVATED COBALT AND CHROMIUM ION LEVELS. RADIOGRAPHIC REVIEWS ALSO RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PINNACLE 100 ACET CUP 54MM CUP LPH DEPUY ORTHOPAEDICS, INC. A3DD71000

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention