FDA Adverse Event Malfunction Summary report: N

ARCHITECT CA 125 II

MDR report key: 2858659 · Received December 5, 2012

Report

Report Number
1415939-2012-02089
Event Type
Malfunction
Date Received
December 5, 2012
Date of Event
October 5, 2012
Report Date
November 12, 2012
Manufacturer
ABBOTT LABORATORIES
Product Code
LTK
PMA / PMN Number
K042731
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ACCURACY TESTING WAS REVIEWED FOR A REPRESENTATIVE LOT OF ARCHITECT CA 125 II (12341M500), SINCE THE LIKELY CAUSE LOT EXPIRED JUNE 22, 2012 AND FURTHER EVALUATION WAS NOT ABLE TO BE PERFORMED. ACCEPTANCE CRITERIA WERE MET, WHICH INDICATES ACCEPTABLE PRODUCT PERFORMANCE. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED FOR THE LIKELY CAUSE LOT 04202M500. DURING THE REVIEW OF THE COMPLAINT TEXT IT WAS NOTED THAT THE REAGENT KIT EXPIRED JUNE 22, 2012 AND THE DISCREPANT RESULTS WERE OBTAINED FOR THE PATIENT ON (B)(6) 2012 BEYOND THE EXPIRATION OF THE KIT. THE ARCHITECT CA 125 II REAGENT PACKAGE INSERT WAS REVIEWED AND WAS FOUND TO ADEQUATELY ADDRESS THE ISSUE. THE INVESTIGATION DID NOT IDENTIFY A MALFUNCTION/DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT ON (B)(6) 2012, A FALSELY ELEVATED ARCHITECT CA 125 RESULT OF 698 U/ML WAS GENERATED ON A PATIENT HAVING A GENERAL PHYSICAL EXAM. THE SAMPLE WAS REPEATED AND A RESULT OF 736 U/ML WAS GENERATED. THE PATIENT RETURNED ON (B)(6) 2012, AND A NEW SAMPLE GENERATED A RESULT OF 25.0 U/ML. A PLASMA SAMPLE FROM (B)(6) 2012, WAS TESTED AND A RESULT OF 614 U/ML WAS GENERATED. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT CA 125 II LTK ABBOTT LABORATORIES 04202M500

Patients

Seq Age Sex Outcome Treatment
1 27 YR ARCHITECT I2000SR ANALYZER 3M74-01| SN (B)(4)| SN (B)(4)| ARCHITECT I2000SR ANALYZER 3M74-01