FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2856261
·
Received October 31, 2012
Report
- Report Number
- 1824206-2012-06995
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 4, 2012
- Report Date
- October 4, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THAT THERE WAS FLUID BEHIND THE GCI AND SHORTED IT OUT. HE REPLACED THE GASKET AND GCI TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATED THAT THE GRAPHICAL CAREGIVER INTERFACE IS NOT FUNCTIONING PROPERLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|