FDA Adverse Event Injury Summary report: N

LEICA M844 F40

MDR report key: 2855734 · Received November 30, 2012

Report

Report Number
3003974370-2012-00014
Event Type
Injury
Date Received
November 30, 2012
Date of Event
November 8, 2012
Report Date
November 8, 2012
Manufacturer
LEICA MICROSYSTEMS (SCHWEIZ) AG
Product Code
EPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS A FINAL REPORT. DURING A SURGICAL PROCEDURE, THE ASSISTANT BINOCULAR TUBE FELL ONTO THE PT'S FOREHEAD. THE PT'S UPPER LEFT EYEBROW WAS INJURED (APPROX 20 MM) AND THE PT RECEIVED MEDICAL INTERVENTION (WOUND SUTURE ABOUT 6 TIMES) AND WAS HOSPITALIZED. THE MFR INVESTIGATED THE COMPLAINT. PHOTOGRAPHIC INSPECTION WAS PERFORMED AND RESULTS SHOWED THAT THE MALFUNCTION COULD BE ATTRIBUTED TO AN INTEROPERABILITY PROBLEM MOST LIKELY CAUSED BY ANOTHER DEVICE. AS DESCRIBED IN THE USER MANUAL (LEICA M844 - LEICA M820 / REF. 10 713 294 / VERSION L, P. 25), THE ASSISTANT BINOCULAR TUBE MUST BE CAREFULLY INSERTED TO THE LEICA DOVE TAIL RING AND THE CLAMPING SCREW MUST BE TIGHTENED. THIS MUST BE DONE DURING PREPARATION AND BEFORE A SURGICAL PROCEDURE. HOWEVER, PHOTO SHOWED THAT THE LEICA ASSISTANT BINOCULAR TUBE, WHICH FELL, WAS ATTACHED TO A SURGICAL MICROSCOPE EYE SAFETY FILTER, WHICH IS A DEVICE THAT IS NOT MANUFACTURED BY LEICA. A PHOTO ALSO INDICATES THAT THE SURGICAL MICROSCOPE EYE SAFETY FILTER IS NOT INDICATED FOR USE WITH THE LEICA (B)(4) SURGICAL MICROSCOPE. IN ADDITION, THE SCREW USED WITH THE SURGICAL MICROSCOPE EYE SAFETY FILTER WAS NOT SUPPLIED OR MANUFACTURED BY LEICA.

Description of Event or Problem · 1

ON (B)(4) 2012, LEICA MICROSYSTEMS RECEIVED A COMPLAINT STATING THAT THE ASSISTANT BINOCULAR TUBE FELL ONTO THE PT'S FOREHEAD DURING A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEICA M844 F40 SURGICAL MICROSCOPE EPT LEICA MICROSYSTEMS (SCHWEIZ) AG M844 F40

Patients

Seq Age Sex Outcome Treatment
1 23 YR Other