FLOGARD
Report
- Report Number
- 1416980-2012-06360
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 13, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. "THE CAUSE WAS IDENTIFIED TO BE THAT THE CONFIGURATION OPTION SETTINGS CHECKSUM WAS NOT ZERO." TO CORRECT THIS CONDITION ALL CONFIGURATION OPTION SETTINGS WERE RESET AS PER SERVICE MANUAL. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED FAILURE CODE 94. THIS WAS NOT REPORTED BY THE CUSTOMER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |