FDA Adverse Event Malfunction Summary report: N

OPTIFLOW TRACHEOSTOMY DIRECT CONNECTION

MDR report key: 2855042 · Received December 3, 2012

Report

Report Number
9611451-2012-00863
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
October 30, 2012
Report Date
November 9, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE OPT570 OPTIFLOW TRACHEOSTOMY DIRECT CONNECTION INTERFACE IS USED TO DELIVER HUMIDIFIED OXYGEN TO PATIENTS VIA TRACHEOSTOMY. THE INTERFACE IS HELD IN PLACE BY A NECK STRAP AND ALSO INCLUDES A LANYARD WHICH IS PLACED AROUND THE PATIENT'S NECK OR ATTACHED TO THE PATIENT'S CLOTHING OR BEDDING TO REMOVE THE LOAD OF THE BREATHING CIRCUIT FROM THE PATIENT'S TRACHEA. METHOD: ONE OF THE THREE COMPLAINT OPT570 INTERFACES WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED THAT THE BREATHABLE FILM OF THE OPT570 TUBING WAS PULLED APART, CONFIRMING THE REPORTED FAULT. A LOT CHECK WAS NOT PERFORMED AS LOT NUMBER WAS NOT PROVIDED. CONCLUSION: THE DAMAGE APPEARED TO BE CAUSED BY THE TUBE BEING PULLED AWAY FROM THE BREATHING CIRCUIT CONNECTOR. ALL OPT570 INTERFACES ARE VISUALLY INSPECTED BEFORE LEAVING THE PRODUCTION LINE AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. FISHER & PAYKEL HEALTHCARE USER INSTRUCTIONS THAT ACCOMPANY THE OPT570 INTERFACE SPECIFY IN THE WARNING SECTION "DO NOT CRUSH OR STRETCH THE TUBE". (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THREE OPT570 OPTIFLOW TRACHEOSTOMY DIRECT CONNECTION INTERFACES WERE "DAMAGED BEHIND CONNECTOR, FRAYING OF THE MATERIAL BETWEEN THE SPINES OF THE TUBE". THIS WAS OBSERVED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPTIFLOW TRACHEOSTOMY DIRECT CONNECTION BZA BTT FISHER & PAYKEL HEALTHCARE LIMITED OPT570 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1