FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2854864 · Received December 3, 2012

Report

Report Number
2531779-2012-14183
Event Type
Malfunction
Date Received
December 3, 2012
Report Date
November 6, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, THE UP ARROW AND CONTRAST BUTTONS HAD INTERMITTENT RESPONSE REQUIRING MULTIPLE PRESSES TO EVOKE A RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2012 REPORTING THAT THE UP ARROW BUTTON WAS INTERMITTENTLY RESPONDING FOR TWO TO THREE MONTHS. THE PATIENT STATED THAT THE BUTTON WAS RESPONDING ABOUT 1 OUT OF 3 PRESSES. THE PATIENT STATED THAT THE BUTTON SYMBOLS WERE STARTING TO WEAR BUT THERE WAS NO NOTED DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP IN A PANTS POCKET AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 22 YR