ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14183
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Report Date
- November 6, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4): ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, THE UP ARROW AND CONTRAST BUTTONS HAD INTERMITTENT RESPONSE REQUIRING MULTIPLE PRESSES TO EVOKE A RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2012 REPORTING THAT THE UP ARROW BUTTON WAS INTERMITTENTLY RESPONDING FOR TWO TO THREE MONTHS. THE PATIENT STATED THAT THE BUTTON WAS RESPONDING ABOUT 1 OUT OF 3 PRESSES. THE PATIENT STATED THAT THE BUTTON SYMBOLS WERE STARTING TO WEAR BUT THERE WAS NO NOTED DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP IN A PANTS POCKET AND DID NOT CLEAN THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |