ENSEAL G2 STRAIGHT JAW
Report
- Report Number
- 3005075853-2012-05432
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 8, 2012
- Report Date
- November 13, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4): ADDED ADDITIONAL INFORMATION. THE INSTRUMENT WAS RECEIVED WITH THE UPPER JAW DETACHED AND THE I-BLADE CRACKED. THE UPPER JAW WAS NOT RETURNED WITH THE INSTRUMENT. THE INSTRUMENT WAS TESTED ELECTRICALLY AND IT ACTIVATED AS EXPECTED WHEN CONNECTED TO THE GENERATOR. DUE TO THE RETURNED CONDITION OF THE INSTRUMENT NOT ALL TESTING COULD BE PERFORMED.THERE IS INSUFFICIENT EVIDENCE RELATED TO WHAT CAUSED THE DAMAGE; A PROBABLE CAUSE OF THE DAMAGE TO THE JAWS/I-BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, WHEN THE DEVICE WAS USED FOR THE THIRD BITE THE TOP JAW CAME OFF. THEY OPENED A SECOND DEVICE, RETRIEVED THE TOP JAW. THE PROCEDURE WAS COMPLETED WITH THE SECOND DEVICE. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENSEAL G2 STRAIGHT JAW | ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | NA | J4C20A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |