FDA Adverse Event Malfunction Summary report: N

PFM PICC II

MDR report key: 2854728 · Received November 29, 2012

Report

Report Number
2032582-2012-00012
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
June 14, 2012
Report Date
November 27, 2012
Manufacturer
PFM MEDICAL, INC.
Product Code
LJS
PMA / PMN Number
K083873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOWING REVIEWS WERE CONDUCTED AND THEIR RESPECTIVE RESULTS WERE FOUND: THERE IS NO IMPACT TO CURRENT INVENTORY AS NO FURTHER SHIPMENTS ARE PLANNED TO BE SENT TO THE COMPLAINANT. A REVIEW OF SIMILAR COMPLAINTS WAS CONDUCTED AS FAR BACK AS 2008 REVEALED THAT NO SIMILAR INCIDENTS OF THIS NATURE HAD BEEN REPORTED TO PFM. SINCE 2011, PFM MANUFACTURED / SOLD 8,164 5F DUAL LUMEN PICC'S. THE RISK OF "CRACKED VALVE" AND "CATHETER LEAK" WAS IDENTIFIED IN THE RISK ANALYSIS. A DHR REVIEW REVEALED THAT THE QUALITY RECORDS FOR LOT# 1111-012 ARE COMPLETE AND IN ORDER. NO NON-CONFORMITIES ARE ASSOCIATED WITH THIS LOT. THE SUSPECTED/FAILED DEVICE WAS NOT RETURNED TO PFM. THEREFORE, WE WERE UNABLE TO TEST THE ACTUAL FAILED DEVICE. INSTEAD, PICCS FROM A SIMILAR LOT WERE TESTED. FOLLOWING 3 DAYS OF SIMULATED USE, NO CRACKING, CRAZING OR LEAKAGE OF FLUIDS WAS OBSERVED. EACH SAMPLE TESTED ALL PASSED; NONE OF THE SAMPLES LEAKED WHEN FLUIDS WERE INFUSED THROUGH THE INJECTION PORT. DUE TO NOT HAVING THE ACTUAL FAILED DEVICE FOR TESTING, AND TESTING ON INVENTORY SAMPLES ALL PASSED, PFM IS UNABLE TO DETERMINE THE ROOT CAUSE OF THE FAILED DEVICES. PFM WILL CONTINUE TO MONITOR FUTURE INCIDENTS OF THIS KIND.

Description of Event or Problem · 1

AT AN UNSPECIFIED DATE/TIME, "THE PROBLEM WAS INITIALLY BROUGHT TO THEIR ATTENTION WHEN THEY WERE TOLD THAT SEVERAL PORTS HAD CRACKED BUT THE INFO WAS NOT MADE AVAILABLE, SO THEY DID NOT FILE A REPORT TO PFM. AS THE PROBLEMS PROGRESSED THEY BEGAN TO PROVIDE NAMES TO IT WAS FINALLY REPORTED TO PFM." ON (B)(4) 2012, FOUR (4) REPORTS WERE PROVIDED TO PFM WITH SIMILAR PROBLEMS AND WE SPLIT THEM INTO FOUR SEPARATE CASES (REF MDR# 203582-2012-00010-13/ PSR00379-382) BECAUSE THE DISTRIBUTOR HAD PROVIDED INFO ON THESE 4 REPORTS ALL ON THE SAME DAY INSTEAD OF SUBMITTING THEM AS INFO BECAME AVAILABLE. THE THIRD CASE INCLUDING THE FOLLOWING COMPLAINT: "PICC PORT LEAKING REQUIRING REPLACEMENT ON (B)(6) 2012."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PFM PICC II PICC LJS PFM MEDICAL, INC. 1111-012

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention