FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2854716 · Received December 3, 2012

Report

Report Number
2531779-2012-14175
Event Type
Malfunction
Date Received
December 3, 2012
Report Date
November 9, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. THE KEYPAD WAS FOUND TO BE INTACT. THE CONTRAST BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING. ALL OTHER BUTTONS WERE FOUND TO BE RESPONDING DURING EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2012, WITH THE FOLLOWING FINDINGS: CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 57 YR