ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-14176
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Report Date
- November 9, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. THE KEYPAD WAS FOUND TO BE INTACT. THE CONTRAST BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING. ALL OTHER KEYPAD BUTTONS WERE FOUND TO BE RESPONDING PROPERLY. UNRELATED TO THE EVENT THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2012 WITH THE FOLLOWING FINDINGS: CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |