ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14153
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Report Date
- November 6, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/06/2012 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE UP AND CONTRAST BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. EVALUATION REVEALED THAT THERE WAS CONTAMINATION UNDER ALL OF THE BUTTON CONTACTS. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE CONTACT BUTTONS. THE UP AND CONTRAST BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |