FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 2854601
·
Received December 3, 2012
Report
- Report Number
- 3015876-2012-00867
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 6, 2012
- Report Date
- November 6, 2012
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K103567
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT FAILED TO BOOT UP PROPERLY BY POWERING ON/OFF ON IT'S OWN (RESETTING) INTERMITTENTLY. PHYSIO REPLACED THE SYSTEM PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.