FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 2854601 · Received December 3, 2012

Report

Report Number
3015876-2012-00867
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 6, 2012
Report Date
November 6, 2012
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K103567
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT FAILED TO BOOT UP PROPERLY BY POWERING ON/OFF ON IT'S OWN (RESETTING) INTERMITTENTLY. PHYSIO REPLACED THE SYSTEM PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.