FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2854461 · Received December 3, 2012

Report

Report Number
3004209178-2012-91063
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 18, 2012
Report Date
November 18, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, PRIME AND DISPLACEMENT TESTS. HOWEVER, THE INSULIN PUMP WAS RECEIVED STUCK IN THE MOTOR ERROR LOOP DURING BOLUS AND BASAL DELIVERY. AN ERROR ALARM WAS CONFIRMED IN THE HISTORY FILE. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HER INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 129MG/DL. TROUBELSHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CALLER TO RUN THE DISPLACEMENT TEST AND THE TEST FAILED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 56 YR