FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2854454 · Received December 3, 2012

Report

Report Number
3004209178-2012-91055
Event Type
Injury
Date Received
December 3, 2012
Date of Event
November 9, 2012
Report Date
November 17, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE DIABETES KETOACIDOSIS AND TO HIGH BLOOD GLUCOSE OF 411MG/DL. THE CALLER STATED THAT SHE IS PREGNANT. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND BOLUS WIZARD SETTINGS WERE CORRECT. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND PASSED. ADVISED THE CUSTOMER TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 30 YR Hospitalization