FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2854441 · Received December 3, 2012

Report

Report Number
3004209178-2012-91039
Event Type
Injury
Date Received
December 3, 2012
Date of Event
November 14, 2012
Report Date
November 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETES KETOACIDOSIS OF 867MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME AND DATE WERE INCORRECT. ASSISTED THE CALLER TO CORRECT THEM. REVIEWED THE BOLUS HISTORY AND FOUND THAT THE CUSTOMER BOLUSED A FEW TIMES ON SOME OF THE DAYS AND ONCE IN ANOTHER DAYS. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND PASSED. ADVISED THE CALLER TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization