FDA Adverse Event Malfunction Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25PA

MDR report key: 2854416 · Received November 29, 2012

Report

Report Number
3003637274-2012-00058
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
October 25, 2012
Report Date
November 2, 2012
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE HANDPIECE DID SHOW THAT THE BEARINGS HAVE BEEN GRITTY AND THE HEAD HAD SEVERAL DENTS CAUSING THE BACKCAP BUTTON TO STICK IN. THIS CAUSES UNUSUAL INNER FRICTION CAUSING THE HEAD TO HEAT UP. THE DENT INDICATES THE HANDPIECE WAS DROPPED OR MISHANDLED, E.G. DURING REPROCESSING. THE CONDITION OF THE BEARING MIGHT BE A NORMAL WEAR ISSUE, BUT IT IS LIKELY THAT DUE TO THE DENT AND THE HIGH TEMPERATURE A STRONGER WEAR TOOK PLACE. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL INSPECTION PRIOR TO EACH TREATMENT WHICH WOULD HENCE BE MANDATORY AND SHOW WHETHER PRODUCT IS RUNNING WITHIN SPECIFICATION. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT ON TOOTH #5 THE HANDPIECE HEATED UP AND CAUSED A BURN ON THE LOWER LIP OF THE PATIENT. SOME OTC ALOE CREAM WAS APPLIED TO THE BURN BUT NO MEDICAL CARE WAS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25PA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 85 YR