FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2854406
·
Received December 3, 2012
Report
- Report Number
- 3004209178-2012-91048
- Event Type
- Injury
- Date Received
- December 3, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 16, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS STUCK IN MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. HOWEVER, THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING TESTING.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 700MG/DL. TROUBLESHOOTING WAS PERFORMED. THE FATHER STATED THAT THE HEALTH CARE PROFESSIONAL BELIEVES THE INSULIN PUMP WAS NOT FUNCTIONING PROPERLY AND REQUESTED THE DEVICE BE REPLACED. THE CALLER ALSO STATED THAT THE DEVICE ALARMED MOTOR ERROR, AND THE BATTERY WAS REMOVED FROM THE INSULIN PUMP. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Hospitalization |