FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 2854325 · Received November 28, 2012

Report

Report Number
2518422-2012-02336
Event Type
Malfunction
Date Received
November 28, 2012
Date of Event
October 30, 2012
Report Date
October 30, 2012
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SVC CTR, "VENTILATOR INOPERATIVE" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR AND SYSTEM BOARD WERE REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A VENTILATOR WAS MAKING NOISE AND ALARMING. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1