FDA Adverse Event Injury Summary report: N

FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER

MDR report key: 2854076 · Received December 3, 2012

Report

Report Number
3005099803-2012-05358
Event Type
Injury
Date Received
December 3, 2012
Date of Event
October 23, 2012
Report Date
November 8, 2012
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
GEX
PMA / PMN Number
K100078
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL ANALYSIS OF THE RETURNED FIBER REVEALED THERE IS NO EXPOSED GLASS TIP REMAINING. THE PATTERN ON THE TIP FACE IS CONSISTENT WITH NORMAL USE INDICATING THE FIBER DEGRADED DURING USE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT LESS THAN FOUR MINUTES INTO THE BLADDER STONE PROCEDURE, THE TIP OF THE FLEXIVA 1000 LASER FIBER BROKE OFF INSIDE THE PATIENT AND WAS RETRIEVED. THE DEVICE USED TO RETRIEVE THE FIBER TIP IS UNKNOWN. LASER TYPE AND LASER SETTINGS ARE UNKNOWN. THE CASE WAS COMPLETED WITH A DIFFERENT DEVICE WITHOUT ANY COMPLICATIONS TO THE PATIENT WHO IS REPORTEDLY 'FINE' POST PROCEDURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT LESS THAN FOUR MINUTES INTO THE BLADDER STONE PROCEDURE, THE TIP OF THE FLEXIVA 1000 LASER FIBER BROKE OFF INSIDE THE PATIENT AND WAS RETRIEVED. THE DEVICE USED TO RETRIEVE THE FIBER TIP IS UNKNOWN. LASER TYPE AND LASER SETTINGS ARE UNKNOWN. THE CASE WAS COMPLETED WITH A DIFFERENT DEVICE WITHOUT ANY COMPLICATIONS TO THE PATIENT WHO IS REPORTEDLY "FINE" POST PROCEDURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER POWERED LASER SURGICAL INSTRUMENT GEX BOSTON SCIENTIFIC - MARLBOROUGH M0068403941 ML00000814

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention