FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC

MDR report key: 2854044 · Received October 31, 2012

Report

Report Number
1824206-2012-06974
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 11, 2012
Report Date
October 11, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN CLEANED THE BOARD AND RECONNECTED ALL CABLING AND RESET THE BED TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THAT LIQUID HAD ENTERED THE RIGHT CAREGIVER PC BOARD CAUSING THE BED TO MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BARIATRIC AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1