FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2853988
·
Received October 24, 2012
Report
- Report Number
- 1811755-2012-03891
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- September 4, 2012
- Report Date
- September 25, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS DISASSEMBLED TO BE VISUALLY EXAMINED. DURING THE VISUAL EXAMINATION, THE DEVICE WAS FOUND TO HAVE CORRODED COMPONENTS INCLUDING THE DRIVESHAFT ASSEMBLY, SPINDLE HOUSING, ROTOR, NOSECONE, AND BEARING STOP. BASED ON THE FINDINGS OF THE VISUAL EXAMINATION, THE OBSERVED HEAT WAS MOST LIKELY DUE TO THE CORROSION ON AND AROUND THE DRIVESHAFT ASSEMBLY.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING A PROCEDURE. A BACK-UP DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT PT OR USER INJURIES, AND THERE WERE NO ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO | 0720708823 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |