FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2853973 · Received October 24, 2012

Report

Report Number
1811755-2012-03875
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
October 2, 2012
Report Date
October 2, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE ROTOR, DRIVESHAFT AND SPINDLE HOUSING WERE FOUND TO BE WORN. THE FRICTION CAUSED BY THE WORN COMPONENTS WAS IDENTIFIED AS A POSSIBLE CAUSE OF OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DRILL WAS OVERHEATING. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK