FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2853947 · Received October 24, 2012

Report

Report Number
1811755-2012-03873
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
October 1, 2012
Report Date
October 2, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE UPPER END OF THE ROTOR WAS FOUND TO BE DAMAGED, AND THE BEARINGS ON THE DRIVESHAFT ASSEMBLY AS WELL AS THE BEARINGS ON THE ROTOR ASSEMBLY WERE NOT TURNING SMOOTHLY, WHICH ARE PROBABLE CAUSES OF THE REPORTED OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK