FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2853947
·
Received October 24, 2012
Report
- Report Number
- 1811755-2012-03873
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- October 1, 2012
- Report Date
- October 2, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVALUATION, THE UPPER END OF THE ROTOR WAS FOUND TO BE DAMAGED, AND THE BEARINGS ON THE DRIVESHAFT ASSEMBLY AS WELL AS THE BEARINGS ON THE ROTOR ASSEMBLY WERE NOT TURNING SMOOTHLY, WHICH ARE PROBABLE CAUSES OF THE REPORTED OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |