FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT

MDR report key: 2853835 · Received December 3, 2012

Report

Report Number
9611451-2012-00881
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 5, 2012
Report Date
November 9, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE EXPIRATORY AND INSPIRATORY LIMBS OF THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. THE HEATER WIRE PINS WERE TESTED TO SEE IF THEY COULD BE CONNECTED TO A HEATER WIRE ADAPTOR. RESULTS: FULL INSERTION OF THE HEATER WIRE ADAPTOR ON THE EXPIRATORY LIMB WAS NOT ACHIEVED AS ONE OF THE HEATER WIRE PINS WAS BENT. NO FAULT WAS FOUND WITH THE INSPIRATORY HEATER WIRE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120713. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE INSPIRATORY HEATER WIRE PINS WERE DAMAGED POST PRODUCTION. IT IS POSSIBLE FOR THE USER TO DAMAGE THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR BENT PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE BENT DURING PRODUCTION OR BY THE END USER. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT RT340 ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE EXPIRATORY HEATER WIRE PINS OF AN RT340 ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT WERE DAMAGED, PREVENTING CONNECTION OF THE BREATHING CIRCUIT TO THE HEATER WIRE ADAPTOR. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE EXPIRATORY HEATER WIRE PINS OF AN RT340 ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT WERE DAMAGED, PREVENTING CONNECTION OF THE BREATHING CIRCUIT TO THE HEATER WIRE ADAPTOR. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT340 120713

Patients

Seq Age Sex Outcome Treatment
1