FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2853770
·
Received December 3, 2012
Report
- Report Number
- 2032227-2012-07919
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 20, 2012
- Report Date
- November 20, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND AND BASIC OCCLUSION TEST. THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM STUCK IN BOLUS DELIVERY LOOP. MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE NOT DETECTED DURING OUR TESTING.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC MINIMED | MMT-523LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |