FDA Adverse Event
Malfunction
Summary report: N
SMALL GRASPING RETRACTOR INSTRUMENT
MDR report key: 2853524
·
Received December 3, 2012
Report
- Report Number
- 2955842-2012-01144
- Event Type
- Malfunction
- Date Received
- December 3, 2012
- Date of Event
- November 2, 2012
- Report Date
- November 2, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING OF THE SMALL GRASPING RETRACTOR INSTRUMENT, THE WIRE ON THE INSTRUMENT WAS OBSERVED TO BE BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SMALL GRASPING RETRACTOR INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420318-02 | M10120528 336 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM AND ACCESSORIES. |