FDA Adverse Event Malfunction Summary report: N

SMALL GRASPING RETRACTOR INSTRUMENT

MDR report key: 2853524 · Received December 3, 2012

Report

Report Number
2955842-2012-01144
Event Type
Malfunction
Date Received
December 3, 2012
Date of Event
November 2, 2012
Report Date
November 2, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE SMALL GRASPING RETRACTOR INSTRUMENT, THE WIRE ON THE INSTRUMENT WAS OBSERVED TO BE BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SMALL GRASPING RETRACTOR INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420318-02 M10120528 336

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM AND ACCESSORIES.