FDA Adverse Event
Summary report: N
RELIANCE SYNERGY WASHER
MDR report key: 2853355
·
Received November 30, 2012
Report
- Report Number
- 9680353-2012-00113
- Date Received
- November 30, 2012
- Date of Event
- November 2, 2012
- Report Date
- November 30, 2012
- Manufacturer
- STERIS CANADA CORPORATION
- Product Code
- LDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A STERIS SERVICE TECHNICIAN INSPECTED THE UNIT AND FOUND WATER TO BE LEAKING FROM THE ORANGE FLEXIBLE HOSE BETWEEN THE DRYER HEATER AND THE DRYING MANIFOLD AND VALVE. THE TECHNICIAN FOUND THAT THE DRYER PISTON HAD FAILED DUE TO THE CENTER PIECE OF THE VALVE BEING BROKEN. THE TECHNICIAN REPAIRED THE UNIT, RAN A TEST CYCLE AND CONFIRMED IT TO BE OPERATIONAL.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM UNDERNEATH THEIR RELIANCE SYNERGY WASHER. THE AMOUNT OF WATER CREATED A PUDDLE LARGER THAN 3'X3'. NO INJURIES, PROCEDURAL DELAYS/CANCELLATIONS OR PROPERTY DAMAGE WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RELIANCE SYNERGY WASHER | WASHER | LDS | STERIS CANADA CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |