FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 2853355 · Received November 30, 2012

Report

Report Number
9680353-2012-00113
Date Received
November 30, 2012
Date of Event
November 2, 2012
Report Date
November 30, 2012
Manufacturer
STERIS CANADA CORPORATION
Product Code
LDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE UNIT AND FOUND WATER TO BE LEAKING FROM THE ORANGE FLEXIBLE HOSE BETWEEN THE DRYER HEATER AND THE DRYING MANIFOLD AND VALVE. THE TECHNICIAN FOUND THAT THE DRYER PISTON HAD FAILED DUE TO THE CENTER PIECE OF THE VALVE BEING BROKEN. THE TECHNICIAN REPAIRED THE UNIT, RAN A TEST CYCLE AND CONFIRMED IT TO BE OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM UNDERNEATH THEIR RELIANCE SYNERGY WASHER. THE AMOUNT OF WATER CREATED A PUDDLE LARGER THAN 3'X3'. NO INJURIES, PROCEDURAL DELAYS/CANCELLATIONS OR PROPERTY DAMAGE WERE REPORTED.